Xantrex Technology XTR 850 Watt Power Supply User Manual


 
Remote Operation
5-30 975-0200-01-01
Figure 5-20 summarizes the Standard Event Status Register.
Figure 5-20
Summary of Standard Event Status Register
Table 5-10
Standard Event Status Register
Bit Bit Weight Bit Name Description
0 1 Operation Complete
(OPC)
Set if *OPC command has been received and all
pending operations have been completed.
1 2 Request Control (RQC) Not implemented. Always set to 0.
2 4 Query Error (QYE) Not implemented.
3 8 Device Dependent Error
(DDE)
Set if there is a device-specific error.
See “Device-Specific Error List” on page B–4
for a list possible error codes and descriptions.
4 16 Execution Error (EXE) Set if a program data element, following a header,
was evaluated by the power supply as outside of its
legal input range, or is otherwise inconsistent with
the power supply’s capabilities. Suggests that a
valid program message could not be properly
executed due to some power supply condition.
See “Execution Error List” on page B–3 for a list of
possible error codes and descriptions.