3 Display and Keyboard
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 3
3.5 Important Functions of the Softkeys
DISPLAY
♦ Change from the single component display to the multi component display.
MEASURE
♦ Change from the main menu to the single component display.
STATUS (see 3.7!)
♦ Change to the status report: Display of the most important parameters and
status informations of the CLD module or of the current channel.
♦ If available, this command is always assigned to the F2 key.
CLD
F1 F2 F3 F4 F5
95.0 ppm
Basic Controls
Measurement range number: 4
Range and functional control: LOCAL
Ranges with valid calibration: 4
Calibration status: READY
If it won´t calibrate...
Status: STANDBY
Ozonator status: On
Ozonator power: Enabled
Range upper limit: 250.0 ppm
Measurement mode: NOx
HOME NO/NOx ZERO SPAN INFO
HOME FACTORS ZERO SPAN INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Zero/span calibration
Measurement range number: 3
Span gas concentration: 100.0 ppm
Sample flow: 1100 ml/min
Raw measurement signal: 521590
Measurement gas: NOx
NO/NOx toggle!
Result...
Calibration adjustment limits: Disabled
Zero gas concentration: 0.0 ppm
Status: READY
Line of Variables
Text Line
Function Line
Menu Line
Function Key Legend