6 - 42
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
Start up analyzer
↓↓
In the menu "Analyzer start up" you will find the actual and target temperatures of the
converter and of the detector block. Besides, you can reboot the analyzer module and re-
initialize the EPROM in this menu:
1) Reboot:
• To restart the analyzer you have to press the F3 key.
• This reboot will start your analyzer in the same way as you do by switching on.
• If you flow the analyzer with zero gas during this time: The NO signal should be "0".
2) Re-initializing of the EPROM:
• If you press the F4 key you will update the EPROM data.
• Caution:
The analyzer will go back to the default values using this function !
After that you have to do a new start up, completely !
HOME ESCAPE REBOOT INIT. INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Analyzer start up
(Press INIT to re-initialize the EEPROM data)
Sensor temperature: 27.0 C
Detector block temperature: 50.3 C
Target temperature: 51.5 C
Converter temperature: 302.0 C
Target temperature: 300.0 C
Analyzer state: STANDBY
6.3.2.9 Diagnostic Menus
Analyzer start up