Emerson NGA2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 9
Main Menu — Expert controls and set up
Analyzer module set up
Calibration parameters
In the menu "Calibration Parameters" you can set up several parameters of the zero and
span calibration.
Set up parameters:
Select the line of variables you want with the -key or the -key.
Select the variable with the ENTER key or the -key.
Select any digit with the -key or the -key and adjust a new value with
the -key or the -key
resp. select the whole parameter with the -key or the -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Line of variables "Calibration adjustment limits:
Enabled: The calibration result will only be accepted within the limit settings. If the
results will be out of these limits, you will find a corresponding error message in the
menu "Analyzer zero" respectively "Analyzer span". Illustrations of these menus you will
find in chapter 4.2.1 p. 4-22 resp. 4.2.2 p. 4-26.
Disabled: The calibration will be possible, even if the calibration results will leave the
allowed limits. Compare with the explanations in chapter 5.1.1 p. 5-6/7.
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Calibration Parameters
Calibration adjustment limits: Disabled
Calibration failure alarm: No
Cal. failure error allowed: 50 %
Calibration time out: 60 s
Zero ranges: TOGETHER
Calibration averaging time: 5 s
Span ranges: SEPARATELY
5.1.2 Analyzer Module Set up
Calibration Parameters