Emerson NGA2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 35
Serial interface setup:
Press the ENTER key or the -key in the line "Serial interface setup..." of the menu
"LOCAL SIO CONFIGURATION SECTIONS" to change to the corresponding submenu:
In the submenu "SERIAL INTERFACE SETUP" you can determine the parameters of the
data transfer between the MLT analyzer or platform and external implements. The setups
in this menu are depending on the configuration of the analyzer resp. platform and the
corresponding implement. The specification of the serial interface is described in another
manual.
Set up parameters:
Select any line of variables with the -key or the -key.
Select the variable with the ENTER key or the -key.
Select any digit with the -key or the -key and adjust a new value with
the -key or the -key
resp. select the whole parameter with the -key or the -key.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
Options:
Baudrate: 300 1200 2400 4800 9600 19200
Data bits: 7 8
Stop bits 1 2
Parity: None Even Odd
Echo mode: Enabled Disabled
Handshake: None Xon/Xoff
Type of installed
serial interface:
RS 232 RS 485_2X RS 485_4X None
EXPERT
CLD
F1 F2 F3 F4 F5
95.0 ppm
- SERIAL INTERFACE SETUP -
Baudrate: 19200
Stop bits: 1
Parity: None
Echo mode: Disabled
Handshake: Xon/Xoff
Data bits: 8
Type of installed serial interface: RS232
INFO