System Calibration
Supplement - 8 NGA 2000 90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
Example for an analyzer-system:
V1 AM1 AM2 AM3
V2 AM4
V3 AM5
V 4 V 5 V 6
Calibration gas mixtures
Appropriate assignment:
AM1 AM2 AM3 AM4 AM5
SAMPLE-GAS V1 V1 V1 V2 V3
ZERO-GAS V4 V4 V5 V5 V6
SPANGAS-1 V5 V5 V4 V6 V5
SPANGAS-2 V5 V5 V4 V6 V5
SPANGAS-3 V6 V5 V4 V4 V4
SPANGAS-4 V6 V5 V4 V4 V4
This assigning procedure is to perform for all analyzer modules which should be included
in system calibration.
To display the assigned valves and purge times for each AM we can push the softkey
„Overview“.
If we want to exclude an AM from SYSCAL we can do this by entering an invalid valve.
We can watch in the display whether an AM is enabled for system calibration.
Attention: Remember to assign a system valve to an output-port!