Emerson NGA2000 Computer Accessories User Manual


 
5 - 32
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Go to the line you want with the -key or the -key.
Select the tag of the channel with the ENTER key or the -key:
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically.
The tag of the channel selected will appear in the line "Source module".
3) Select any signal:
Change to the line "Choose signal" with the -key.
Press the ENTER key or the -key to change to the submenu "SIGNALS":
Change to the line you want with the -key or the -key.
You can change with the F3 key (>>>) to further menu pages containing other available
signals:
t
90
time
Output delay time
Calibration status
Linearizer
Measurement mode
Operational state
Health
Interference
Validity
Raw signal
Primary Variable
Press the ENTER key or the -key to select the signal:
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically.
The signal selected will appear in the line "Signal name".
4) Determination of the output signal values:
You can determine the signal value for the 0% output and the 100% output in the lines
"Signal value for 0% output" or "Signal value for 100% output". So you have the possibility
to zoom out a certain part of the whole range.
MENU BACK
>>> <<<
INFO
CLD
F1 F2 F4 F5
95.0 ppm
- SIGNALS -
Sample flow:
Sample press.:
Ozonator:
Converter temp.:
Block temperature:
NO/NOx:
Noise level:
Detector temperature:
F3