Emerson NGA2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 27
Menu "Pressure Limits":
Menu "Temperature Limits":
Set up parameters:
Select any line of variables with the -key or the -key.
Select the variable with the ENTER key or the -key.
Select any digit with the -key or the -key and adjust a new value with
the -key or the -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Options:
Pressures: 0 to 3000 hPa
Temperatures: 0 to 500 °C for converter; 0 to 75 °C for sensor and block
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Pressure Limits
Sample capillary upper limit: 490.0 hPa
Ozone supply upper limit: 1050 hPa
Ozone supply lower limit: 700.0 hPa
Sample capillary lower limit: 50.0 hPa
Barometric pressure: 1013 hPa
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Temperature limits
Converter upper limit: 500.0 C
Sensor upper limit: 50.0 C
Block upper limit: 55.0 C
Converter lower limit: 150.0 C
Block lower limit: 45.0 C
Sensor lower limit: 0.0 C