Emerson NGA2000 Computer Accessories User Manual


 
5 - 40
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Line of variables "Invert output":
Select "Enabled" in the line "Invert output" if you want to invert the measurement signal at
the actual digital output, otherwise select "Disabled". The output inversion can be neces-
sary for instance, if you want to combine a failure with an alarm control.
The last five lines of variables in the menu "DIO-module outputs" only display some
adjustments of the DIO output selected! They are not available:
"Module status:" displays the status of the current DIO board.
"Slot ID:" displays the tag number of the DIO board selected.
"Signal name:" displays the signal selected for the current DIO output.
"Signal level:" displays the status of the signal selected.
"Signal comes from:" displays the source module selected for the current DIO output.
Configuration of the DIO module inputs:
Press the ENTER key or the -key in the line "Inputs..." to change to the corresponding
submenu:
In the menu "DIO-MODULE INPUTS" you have to determine the configuration of all of the
eight DIO module inputs!
DIO input configuration setup:
1) Select the input number:
Press the ENTER key or the -key in the line "Inputnumber" and adjust an input
number (1, ..., 8) with the -key or the -key.
Confirm the number selected with the ENTER key.
2) Choose the module:
Change with the -key to the line "Choose module...".
Change to the submenu "ANALYZER MODULES" with the ENTER key or the -key.
All modules joint to the platform or to the analyzer are available.
EXPERT
>>>
INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
- DIO-MODULE INPUTS -
Choose module...
Choose function...
Slot ID: 1
Signal level: Off
Signal comes from: ????
Inputnumber: 1
Signal name: ????