Emerson NGA2000 Computer Accessories User Manual


 
4 - 26
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
HOME ESCAPE SPAN INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Analyzer span
Are you sure?
You must have span gas flowing through the analyzer.
This control does NOT control any auto-calibration
module bound to this analyzer!
If you are sure, press SPAN again now.
Calibration status: READY
Error message for last span: CAL OK
Press the left arrow key when you are done.
ABORT
CLD
F1 F2 F3 F4 F5
95.0 ppm
Analyzer span
Are you sure?
You must have span gas flowing through the analyzer.
This control does NOT control any auto-calibration
module bound to this analyzer!
If you are sure, press SPAN again now.
Calibration status: SPANNING - WAIT
Error message for last span: CAL OK
Press the left arrow key when you are done.
HOME ESCAPE SPAN INFO
CLD
F1 F2 F3 F4 F5
225.0 ppm
Analyzer span
Are you sure?
You must have span gas flowing through the analyzer.
This control does NOT control any auto-calibration
module bound to this analyzer!
If you are sure, press SPAN again now.
Calibration status: READY
Error message for last span: CAL OK
Press the left arrow key when you are done.
Start Spanning
++ Pay attention to
the note at p. 4-27 !
Press
F4
Caution:
You have to
flow enough span gas
through the analyzer
look for a stable signal
before starting the span gas
calibration.
Note:
You can go back to the
menu "Basic Controls"
with the F2 key.
Span gas calibration
is running
Note:
You can abort the
procedure at any time
with the F2 key.
Change to the
main menu
Press
F1
Alternative:
If you want to continue with
further setups of the basic
controls, you can go back to
the corresponding menu with
the F2 key or the -key.