Honeywell 2700 Scanner User Manual


 
6 - 11
6. Remove all the weight from the scale platter and verify the display reads:
0.00 lbs
or
0.00kg
Shift Test
Shift Test Zones
1. Ensure there is no load on the scale platter and verify the remote display reads:
0.00 lbs
or
0.00kg
2. Place a 15 lb. or 7.5kg weight on the scale platter in the center of Shift Test Zone A (see diagram) and verify the remote
display reads between:
14.99 and 15.01 lbs
or
7.495 and 7.505kg
3. Remove the 15 lb. or 7.5kg weight and verify the display reads:
0.00 lbs
or
0.00kg
4. Repeat steps 2 and 3 for each of the remaining Shift Test Zones (B, C, and D).
5. When all weight has been removed, verify that the remote display reads:
0.00 lbs
or
0.00kg
Decreasing Load Test
1. Ensure there is no load on the scale platter and verify the remote display reads:
0.00 lbs
or
0.00kg
2. Place a 30 lb. or 15kg load on the center of the scale platter. Verify the remote display reads between:
29.99 and 30.01 lbs
or
14.995 and 15.005kg
3. Remove 10 lbs. or 7.5kg from the platter and center the remaining weight. Verify the remote display reads between:
19.99 and 20.01 lbs
B
A
C
D