Maintenance
3. Open the rack locks. Lift the racks SB1 to SB4 and let the excess chemical
solution drip from the racks.
White Arrow: Hinged holding plate
on the CD and BX racks
4. Lift the BX rack far enough so that the hinged holding plate can rest on the metal
bracket to let the chemical solution drip from the rack.
5. Remove the splash guard from the CD rack. Then remove the CD rack in the
same way as the BX rack. (See Step 4.)
6. Remove the dripped racks, put them in the tray supplied as an accessory and
take them to a lab basin.
7. In the lab basin, rinse the racks first in cold water, then in warm water (maximum
40 °C).
KODAK PROFESSIONAL SRP 30 Laser Printer 8-41