Agilent Technologies 01664-97005 Switch User Manual


 
To test the disk drive voltages
Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and
assemblies.
WARNING
This procedure is to be performed by service-trained personnel aware of the hazards
involved, such as fire and electrical shock.
Equipment Required
Equipment Critical Specification Recommended
Model/Part
Digitizing Oscilloscope > 100 MHz Bandwidth Agilent 54600A
1 Turn off the instrument, then remove the power cable. Remove the instrument
cover and the disk drive.
2 Reconnect the disk drive cable to the rear of the disk drive. Turn the disk drive over
so that the solder connections of the cable socket are accessible.
3 Connect the power cable, insert the operating system disk, then turn on the
instrument.
4 Insert the disk that contains the functional performance verification software and
enter the test operating system.
5 In the Sys PV menu select the Disk test. Insert a disk that has enough available
bytes to run the test in the disk drive, then select Run-Repetitive.
Troubleshooting
To test the disk drive voltages
5–25