Agilent Technologies 01664-97005 Switch User Manual


 
To remove and replace the front panel and keyboard
1 Using previous procedures, remove the following assemblies:
Handle
Rear Feet
Cover
Disk Drive
Power Supply
Main Circuit board
2 Remove the four screws connecting the front panel.
3 Slide the front panel assembly out the front of the instrument.
4 Slide the spacers out the front of the instrument to remove them.
Be careful not to lose the copper ground strips on the spacer.
When installing the spacer, insert the pins of the spacer in the appropriate holes in the
chassis. You can hold the spacer in place while installing the front panel by holding it with
your finger through the disk drive mounting slot in the chassis.
5
Remove the RPG knob by pulling the knob off the RPG shaft.
6 Disassemble the front panel assembly by lifting the keyboard circuit board away
from the front panel.
7 Lift the elastomeric keypad out of the front panel.
8 Reverse this procedure to assemble and install the front panel assembly.
When assembling the front panel, check that the holes in the elastomeric keypad and the
keyboard circuit board align with the pins on the front panel.
To remove and replace the intensity adjustment
1 Using previous procedures, remove the following assemblies:
Handle
Rear Feet
Cover
Rear Panel
2 Disconnect the intensity adjustment cable from the CRT driver board.
3 Using a nut driver, remove the hex nut attaching the intensity adjustment to the
chassis.
4 Slide the intensity adjustment assembly toward the front and up out of the
instrument.
5 Reverse this procedure to install the intensity adjustment.
When installing the intensity adjustment, check that the keying tab on the adjustment aligns
with the keying hole in the chassis.
Replacing Assemblies
To remove and replace the front panel and keyboard
6–10