Acquire the data
1 Enable the pulse generator channel 1 output (with the LED off).
2 Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic
analyzer to acquire data for at least 100 valid runs as indicated in the pattern
statistics field.
3 When the logic analyzer has acquired at least 100 valid runs, touch Stop. The
Min X-O field in the logic analyzer Pattern Statistics menu should read
94.99–95.00 µs. The Max X-O field should read 95.0095.01 µs. The Avg X-O field
should read 94.99–95.01 µs. Record the results in the performance test record.
Testing Performance
To test the time interval accuracy