9 Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or 100
ps.
a In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set
Marker 1 at 1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at
1.3000 V.
b In the oscilloscope Delta T menu, select Start On Pos Edge 1. Select Stop on Neg
Edge 1.
c Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or 100 ps.
10
Select the clocks to be tested.
a
Select the clock field to be tested, then select J + K as the clock edges.
b Select Done to exit the Master Clock menu.
Testing Performance
To test the multiple-clock, multiple-edge, state acquisition
3–43