Agilent Technologies 01664-97005 Switch User Manual


 
c In the oscilloscope Measure menu, select Measure Chan 2, then select Period. If the
period is more than or equal to 20.000 ns, go to step 4. If the period is less than 20.000
ns but greater than 19.75 ns, go to step 3.
d In the oscilloscope Timebase menu, add 10 ns to the delay.
e In the oscilloscope Measure menu, select Period. If the period is more than or equal to
20.000 ns, decrease the pulse generator Chan 2 Doub in 100-ps increments until one of
the two periods measured is less than 20.000 ns but greater than 19.75 ns.
3
Check the data pulse width. Using the oscilloscope, verify that the data pulse width
is 3.50 ns, +0 ps or100 ps.
a Enable the pulse generator channel 1 and channel 2 outputs. Leave channel 2 output
disabled.
b In the oscilloscope Timebase menu, select Sweep Speed 1.00 ns/div.
c Select Delay. Using the oscilloscope knob, position the data waveform so that the
waveform is centered on the screen.
d In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker
1 at 1.3000 V. Set the Marker 2 Position to Chan 1, then set Marker 2 at 1.3000 V.
e In the oscilloscope Delta T menu, select Start On Pos Edge 1. Select Stop on Neg
Edge 1.
f Select Precision Edge Find.
g If the pulse width is outside the limits, adjust the pulse generator channel 1 width and
select the oscilloscope Precision Edge Find until the pulse width is within limits.
Testing Performance
To test the single-clock, single-edge, state acquisition
3–28