Connect the logic analyzer
11 Using the 6-by-2 test connectors, connect the logic analyzer clock and data channels
listed in the following table to the pulse generator. Install a BNC cable between the
pulse generator channel 2 output and the 6x2 test connector with the logic analyzer
clock leads.
2 Using SMA cables, connect channel 1, channel 2, and trigger of the oscilloscope to
the pulse generator.
Connect the 1664A Logic Analyzer to the Pulse Generator
Testing
Combination
Connect to
Agilent 8131A
Channel 1 Output
Connect to
Agilent 8131A
Channel 1
Output
Connect to
Agilent 8131A
Channel 2 Output
1 Pod 1, channel 3
Pod 2, channel 3
Pod 1, channel 11
Pod 2, channel 11
J-clock
3 Activate the data channels that are connected according to the previous tables.
a Press the Format key.
b Select the field showing the channel assignments for one of the pods being tested.
Press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.
Testing Performance
To test the single-clock, multiple-edge, state acquisition
3–48