Agilent Technologies 01664-97005 Switch User Manual


 
To remove and replace the power supply
1 Using previous procedures, remove the following assemblies:
Handle
Rear Feet
Cover
Disk Drive
WARNING
Hazardous voltages exist on the power supply. To avoid electrical shock, disconnect the
power from the instrument before performing the following procedures. After disconnecting
the power, wait at least three minutes for the capacitors to discharge before servicing the
instrument.
2
Lift the PCB locking pins out of the chassis.
3 Slide the power supply out far enough to reach the power supply cables, then
disconnect them from the power supply.
4 Slide the power supply the rest of the way out the side of the instrument.
5 Reverse this procedure to install the power supply.
Check that the following assemblies are properly installed before installing the power supply:
Monitor
Front Panel
Switch Actuator
Main Circuit Board
To remove and replace the Main Circuit board
1 Using previous procedures, remove the following assemblies:
Handle
Rear Panel
Cover
Disk Drive
Power Supply
2 Disconnect the GPIB and RS-232C cables from the Main Circuit board.
3 Disconnect the front panel cable from the Main Circuit board.
4 Disconnect the monitor board cable from the Main Circuit board.
5 Slide the board out the rear of the instrument.
6 Reverse this procedure to install the Main Circuit board.
Check that the following assemblies are properly installed before installing the Main Circuit
board:
Monitor
Front Panel
Switch Actuator
Replacing Assemblies
To remove and replace the power supply
6–7