c Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or −100 ps.
4
Select the clocks to be tested.
a
Select the clock field to be tested and then select J↑ + K↑ as the clock edges.
b Select Done to exit the Master Clock menu.
Testing Performance
To test the multiple-clock, multiple-edge, state acquisition
3–41