90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
6 - 13
♦ Change to the line you want with the ↑↑ -key or the ↓↓ -key.
If there exist more than eight unbound I/O modules (e.g. Autocal I/O) you can change to
another menu page with the F3 key (MORE).
♦ Press the ENTER key or the →→ -key to select the tag of the I/O module you want:
The display will go back to the menu "Module Binding" automatically.
The tag of the I/O module selected will appear in the line "Proposed bind".
3) Binding:
♦ Press the ENTER key or the →→ -key in the line "Bind selections!" or press the F3 key
(BIND) to bind the modules selected:
The control module will change to the initializing mode and the binding will be done,
automatically.
♦ Repeat the steps 1) to 3) to bind further modules you want.
4) Check bindings:
♦ Press the ENTER key or the →→ -key in the line "View bindings..." to change to the
submenu "Modules Bound" where you will find all I/O modules bound to the channel
selected:
HOME ESCAPE MORE BACK INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Select IO modules
Select the modules you wish to bind to the current
I/O module with 3 alarms
analyzer.
Autocal I/O
Syscal I/O
HOME ESCAPE NEXT INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Modules Bound
Auxiliary module: Syscal I/O
Auxiliary module: NONE
Auxiliary module: NONE
Auxiliary module: NONE
Auxiliary module: NONE
Proposed new bind: NONE
Analyzer module selected: CLD
Auxiliary module: NONE