f) Set up further correction points:
You can repeat the steps c) to e) to determine maximum two further correction points.
Select "Point 2" resp. "Point 3" in the line "Point being measured". Set up the value
you want in the line "Point 2 gas concentration" resp. "Point 3 gas concentration".
Measure the gas concentrations and set up the correction with the F3 key.
g) Enable the correction:
Select "ENABLED" in the line "Correction": The midpoint correction will be done auto-
matically for the range selected with the determined values.
The midpoint correction can be done for each range of the CLD analyzer module.