Emerson NGA 2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 21
6) Correction of the linearization polynom, if necessary:
You can optimize the calculated linearization with maximum three additional points.
Press the ENTER key or the -key in the line "Midpoint correction set up..." of the
menu "Linearization functions" to change to the corresponding submenu:
a) Range selecting:
Range 1 will be selected automatically after entering the menu. If you want to select
another range to correct, press the F4 key (RANGE 2/3/4). The range selected will be
displayed in the headline of the menu.
b) Disable the correction:
You have to disable the correction mode before calculating the correction data.
Otherwise, you will have interferences during the calculation with existing values.
Press the ENTER key or the -key in the line "Correction" to select the parameter.
Set up "DISABLED" with the -key or the -key and confirm it with the ENTER key.
c) Measured point selecting:
Go to the line "Point being measured" with the -key.
Press the ENTER key or the -key to select the parameter. Select "Point 1" with the
the -key or the -key and confirm it with the ENTER key.
d) Set up the gas concentration for point 1:
Go to the line "Point 1 gas concentration" with the -key.
Press the ENTER key or the -key to select the parameter.
Select any digit with the -key or -key and set up the value you want with the the
-key or the -key. If necessary, change the sign with the F4 key ("+/-").
Confirm your set up with the ENTER key.
e) Measuring of the gas concentrations:
Flow the analyzer with the gas concentration set up of the line "Point 1 gas
concentration" and measure the raw value.
After the signal will be stable press the F3 key (SET): The current signal will be
displayed in the line "Point 1 reading". But, the analyzer will equate the raw value with
the desired value of this point automatically.
HOME ESCAPE SET RANGE 4 INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Midpoint correction set up
Correction: DISABLED
Point 1 gas concentration: -1.00 ppm
Point 2 gas concentration: -1.00 ppm
Point 3 gas concentration: -1.00 ppm
Point 1 reading: -1.00 ppm
Point 2 reading: -1.00 ppm
Span gas value: 0.00 ppm
Analyzer functioning: READY
Point being measured: Point 3
Point 3 reading: -1.00 ppm
Range 3