Saving the factors:
You can store the actual zero offset and span factor with the F2 key (STORE) of the
menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu
page (see next illustration).
You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with
the F4 key (HISTORY):
You can change the stored settings against the manufacturer's settings with the F3 key
(RSTR MN: Restore the manufacturer's values). The "Zero offset" and the "Span factor"
of the first menu page will change according to the setups of the manufacturer's
settings.
You can change the manufacturer's settings against the stored settings with the F4 key
(RSTR ST: Restore the "stored" values). The "Zero offset" and the "Span factor" of the
first menu page will change according to the setups of the stored settings.
HOME NEXT RSTR MN RSTR ST INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Range 3 Factors
Zero offset: 0.000000
Zero offset: 0.000000
Span factor: 0.000000
Span factor: 0.000000
Manufacturer´s settings.
Stored settings