5 - 34
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
7) Fine adjustment of the analog output:
If you press the F4 key in the menu "ANALOG OUTPUT SETUP" you can change to the
submenu "OUTPUT FINE ADJUSTMENT", where you can do the fine adjustments of the
analog output:
In the line "Output number" the output number selected in the menu "ANALOG OUTPUT
SETUP" will appear. If you want you can change the number in the submenu "OUTPUT
FINE ADJUSTMENT":
Press the ENTER key or the -key to select the output number and adjust the value
you want with the -key or the -key. Options: 1, ..., 8.
Confirm the number selected with the ENTER key.
Set up fine adjustment of the analog output:
Press the ENTER key or the -key in the line "Operation mode" to select the
parameter and adjust one of the following parameters with the -key or the -key:
Normal: The absolute measurement signal will be sent to the analog output.
Adjust 0V: Coordination between the display and the analog output for 0V with fine
adjustment 0%.
Adjust 10V: Coordination between the display and the analog output for 10V with fine
adjustment 100%.
Confirm the new parameter with the ENTER key.
Press the ENTER key or the -key in the line "Fine adjustments for 0% output" or
"Fine adjustment for 100% output" to select the corresponding value.
Adjust the value you want with the -key or the -key and confirm it with the ENTER
key. Options: 3500 - 4800 for 0% and 750 - 900 for 100%.
The last three lines of variables in the menu "Analog output setup" only display some
values of the analog output! They are not available.
EXPERT BACK
CLD
F1 F2 F3 F4 F5
95.0 ppm
- OUTPUT FINE ADJUSTMENT -
Operation mode: Normal
Fine adjustment for 100% output: 799
Fine adjustment for 0% output: 4097
Output number: 1
INFO