4.1.5 Measurement - Enable/Disable Ozonator Power
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 15
DISPLAY STATUS MENU NEXT INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1 F2 F3 F4 F5
0 250
95.00 ppm NOx
Range: 4
MEASURE STATUS NEXT LOCK INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Main Menu
Basic controls...
Expert controls and set up...
Technical level configuration...
(Operational configuration)
(Diagnostic and manufacturing/service)
HOME NO/NOx ZERO SPAN INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Basic Controls
Measurement range number: 4
Range and functional control: LOCAL
Ranges with valid calibration: 4
Calibration status: READY
If it won´t calibrate...
Status: STANDBY
Ozonator status: On
Ozonator power: Enabled
Range upper limit: 250.0 ppm
Measurement mode: NOx
⇒ ⇒ Change to the
main menu
Press
F3
or
⇒ ⇒ Change to the
basic controls
Press
or
⇒ ⇒ Change to the line
"Ozonator power"
Press or
as often as necessary
until the line "Ozonator
power" will appear white
on black.