Emerson NGA 2000 Computer Accessories User Manual


 
4.2.1 Calibration - Zeroing
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 23
MEASURE STATUS NEXT LOCK INFO
CLD
F1 F2 F3 F4 F5
0.017 ppm
Main Menu
Basic controls...
Expert controls and set up...
Technical level configuration...
(Operational configuration)
(Diagnostic and manufacturing/service)
DISPLAY STATUS MENU NEXT INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1 F2 F3 F4 F5
0.017 ppm NOx
Range: 4
0 250
Change to the
single component
display
Press
F1
Single component
display will appear
after the zero gas
calibration
Notes for zero gas calibration:
You can calibrate all ranges together or each range separately
with zero gas.
You have to set up the parameter you want in the menu
"Calibration Parameters" (see chapter 5.1.2 S. 5-9).