Emerson NGA 2000 Computer Accessories User Manual


 
5 - 24
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Line of variables "Ozonator power":
Options: Enabled/Disabled
The setup of this line will change the displayed message of the line "Ozonator status".
That line is not available. The ozonator power setup is also possible in the menu "Basic
Controls" (see chapter 4.1.5 p. 4-15).
Lines of variables "First/Second/Third/Fourth line's parameter:
In these lines you can set up which of the following parameters shall be displayed in each
of the last four lines of the single component display:
Sample pressure (bargraph with measured value)
Sample flow (bargraph with measured value)
Raw signal (bargraph with measured value)
Validity (report: valid/invalid)
Interference (report: On/Off)
Health (report: Normal/...)
Operational state (report: STANDBY/...)
Measurement mode (report: 1/...)
Linearizer (report: Enabled/Disabled)
Calibration status (report: READY/...)
Output delay time (bargraph with value selected)
t
90
time (bargraph with value selected)
Noise level (of concentration)
NO/NO
x
(Display: NO/NOx)
Detector temperature (bargraph with measured value)
Block temperature (bargraph with measured value)
Converter temperature (bargraph with measured value)
Ozonator (report: OFF - ENABLED/OFF - DISABLED)
Menu line "Linearizer parameters...":
Via this line you can change to the submenu "Linearity coefficients", where you can set up
the coefficients of the linearization curves manual. You can also reach the menu "Linearity
coefficients" via the menu "Linearization parameters". These menus are described
detailed in chapter 5.1.4 p. 5-12/13.
Ramifications via function keys:
With the F3 key (NEXT) of the menu "Analyzer Parameter List" you can change to further
menu pages, where you can set up the following concentration and calibration parameters:
Control mode
Output delay time
Begin and end of all four ranges
t
90
time (response time) for all ranges
Linearizer enabling/disabling for each range
Calibration averaging time