Emerson NGA 2000 Computer Accessories User Manual


 
4 Basic Controls
90003479(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 1
In the chapter "Basic Controls" you will find the most important functions to set up your
CLD analyzer module via an NGA front panel:
1) Measurement p. 4 - 3pp (Chapter 4.1.1 and 4.1.2, p. 4-3 to 4-6, are only important
for systems with several channels.)
2) Calibration p. 4 - 21pp
All steps are figured with detailed illustrations and operation instructions. In the left column
you can see display and keyboard of the NGA front panel. In the right column you can
read the instructions and notes. All instructions will begin with any single component
display and will end with the corresponding single component display after the setups are
done. The keys you have to press are illustrated in black. The instructions are completed
with notes and further informations.
Example: You want to change from the single component display of NO
x
to the main menu
Picture one shows the starting situation: single component display of NO
x
.
If you press the F3 key you will change to menu page shown in picture 2: Main Menu.
Left column: Right column:
LCD and keyboard Instructions and notes
DISPLAY STATUS MENU NEXT INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1 F2 F3 F4 F5
0 250
95.00 ppm NOx
Range: 4
MEASURE STATUS NEXT LOCK INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Main Menu
Basic controls...
Expert controls and set up...
Technical level configuration...
(Operational configuration)
(Diagnostic and manufacturing/service)
Change to the
Main Menu
Press
F3
or
Next instruction
or step,
here e.g.:
Change to the
Basic Controls:
Press
or