Emerson NGA 2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
6 - 31
Set up parameters:
Press the ENTER key or the -key in the line "Zero offset" to select the parameter and
set up "32700" resp. "524000" with the -key or the -key. You may select any digit
with the -key or the -key.
Note: The zero offset is about 32000 counts for the CLD software revision 2.2.1 (16 bit
D/A transformer) and about 520000 counts for the revision 2.3 (20 bit D/A transformer).
Confirm your setup with the ENTER key.
Go to the line "Span factor" with the -key.
Press the ENTER key or the -key to select the parameter and set up "0.00015" with
the arrow keys.
Confirm your setup with the ENTER key.
Notes:
You have to set up the factors for all ranges.
With the parameter "Zero offset" you will fix the zero value. The changing of the zero
offset will influence the signal immediately (look at the headline).
With the parameter "Span factor" you will fix the signal at the span gas setpoint. The
changing of the span factor will influence the signal immediately (look at the headline).
Conditions for the manual input of factors:
"Disabled" has to be set up in the line "Calibration adjustment limits" of the menu
"Zero/span calibration". Look at page 5-5 of chapter 5.1.1.
"SEPARATELY" has to be set up in the lines "Zero ranges" and "Span ranges" of the
menu "Calibration Parameters" (see 5.1.2, p. 5-9).
Attention:
After the calibration with factors you have to calibrate with zero gas and span gas in the
usual way!
Saving the factors:
You can store the actual zero offset and span factor with the F2 key (STORE) of the
menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu
page (see illustration on the next page).
You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with
the F4 key (HISTORY):