Emerson NGA 2000 Computer Accessories User Manual


 
5 - 50
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
System set up
Display resolution
In the menu "Display resolution" you can determine the number of digits for the display of
the measured values. These numbers do not alter the accuracy of measurement. The
adjustments are not valid for the analyzer module!
Set up parameters:
Select one of the two lines of variables with the -key or the -key.
Select the number with the ENTER key or the -key.
Select the digit with the -key or the -key.
Options: Maximum total number of digits: 3, 4, 5, 6
Digits after the decimal point: 0, 1, 2, 3.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
Example:
Measured number: 1.2345
Select 3 for total number of digits and 2 for digits after decimal point.
The screen will display: 1.23
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Display resolution
Maximum total number of digits: 4
Digits after the decimal point: 2
5.4.2 System Set Up
Display resolution