Emerson NGA 2000 Computer Accessories User Manual


 
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 31
Analog Output Setup:
Press the ENTER key or the -key in the line "Analog output setup..." to change to the
corresponding submenu:
Configuration of the analog output:
1) Select the analog output number:
You can set up all parameters in the menu "ANALOG OUTPUT SETUP" separately for
each of the available analog outputs. The number of outputs depends on the hardware
configuration of the SIO module (minimum: 2, maximum: 8).
Press the ENTER key or the -key to select the output number and adjust the value
you want with the -key or the -key.
Confirm the number selected with the ENTER key.
2) Choose the signal source module:
Change with the -key to the line "Choose signal source module..."
Change to the submenu "ANALYZER MODULES" with the ENTER key or the -key.
The tags of all analyzer modules combined with the platform or the MLT analyzer will
appear:
EXPERT ADJUST INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
- ANALOG OUTPUT SETUP -
Output number: 1
Choose signal...
Signal value for 0% output: 0.00 ppm
Signal value for 100% output: 2500.00 ppm
Output current range: 0...20 mA
Hold output during calibration? No
Current signal value: 95.00 ppm
Source module: CLD
Choose signal source module...
Signal name: NO/NOx
MENU BACK
>>> <<<
CLD
F1 F2 F3 F4 F5
95.0 ppm
- ANALYZER MODULES -
CLD