Main Menu — Technical level configuration
System set up
System Reset
You can start the following functions in the menu "System Reset":
1) System Reset:
The MLT analyzer or the platform will change to the initializing mode automatically, if
you press the ENTER key or the -key in the line "System reset!":
This reboot is analogous to the "switching on" of the control module.
2) Re-initialize network:
Change to the line "Re-initialize network!" with the -key or the -key.
Press the ENTER key or the -key to start the function.
Caution: With the function "Re-initialize network!", you will delete all bindings between
analyzer modules and I/O modules and all configuration data of the SIO and DIO modules
will disappear!
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
System Reset
Are you sure?
Re-initializing will destroy all the binds!!!
Re-initialize network!
System reset!
6.1.5 System set up
System Reset