Emerson NGA 2000 Computer Accessories User Manual


 
5 - 18
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
Analyzer module set up
Gas measurement parameters
Linearization functions
From the menu "Linearization functions" you can change to several submenus, where you
can linearize the ranges and optimize the linearization functions:
Change to the line you want with the -key or the -key.
Note:
Normally, the CLD is working linearly in the whole range. Therefore you do not have
to linearize generally ! If you will find deviations from the linearization, you should control
the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for
deviations will be a defect in the ozone supply.
Polynomial set up:
Press the ENTER key or the -key in the line "Polynomial set up..." to change to the
corresponding menu:
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Linearization functions
Polynomial set up...
Use the polynomial set up to generate a linearizing
polynomial from up to 20 gases. With more than 6 gases
it will produce a fourth order polynomial linearizer.
correction, to bring up to three points precisely onto
the curve.
Midpoint correction set up...
Use the midpoint correction for a piecewise-linear final
5.1.8 Analyzer Module Set up
Linearization Functions