Emerson NGA 2000 Computer Accessories User Manual


 
6 - 38
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
Diagnostic menus
Analyzer module diagnostics
Miscellaneous control parameters
In the menu "Miscellaneous control parameters" you will find some converter parameters
and detector parameters. Besides, you can switch on or off the ozonator in this menu..
The set up of the ozonator power is also possible in the menu "Basic Controls" (chapter
4.1.5 p. 4-16) or in the menu "Expert controls" (chapter 5.5 p. 5-54) or in the menu
"Analyzer Parameter List" (chapter 5.1.9 p. 5-24).
Set up parameters:
Press the ENTER key or the -key to select the parameter.
Select "Enabled" or "Disabled" with the the -key or the -key:
The report in the line "Ozonator status" will change according to the selection.
Confirm the new setup with the ENTER key or
cancel and go back to the last one with the F2 key.
HOME ESCAPE INFO
CLD
F1 F2 F3 F4 F5
95.0 ppm
Miscellaneous control parameters
Ozonator status: ON
Peltier device voltage: Normal
Converter current: -0.013 A
Block heater current: -0.027 A
Converter heater duty cycle: 0.100 %
Detector block duty cycle: 0.000 %
Fluistor duty cycle: 0.000 %
Peltier control setting: 2992
Ozonator power: Enabled
Fan duty cycle: 99.00 %
6.3.2.6 Diagnostic Menus
Miscellaneous control parameters