Honeywell 9782 Series Network Card User Manual


 
Conductivity/Resistivity Analyzer/Controller
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/9910-6
10.3 Specifying a Tagname or Other Display String
Introduction
The real-time displays of process values show the instrument’s tagname (or other configurable
fixed nineteen-character string) at the top of the screen.
Procedure
Use the procedure in Table 10-2 to specify the text to be displayed on the alarm stripe when no
alarms or other messages are active.
Table 10-2 Procedure for Specifying a Tagname or Other Display String
Step Screen Action
1
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT
Go to MAINTENANCE | INSTRUMENT SETUP |
INSTRUMENT TAGGING.
MAINTENANCE
OFFLINE FUNCTIONS
INSTRUMENT SETUP
LEAD RESISTANCE COMP
DIAGNOSTICS
| PREV | NEXT
INSTRUMENT SETUP
INSTRUMENT TAGGING
PRODUCT INFO
SCREEN CONTRAST
SECURITY
RESET UNIT
DIAGNOSTICS
| PREV | NEXT
2
UPDATE TAG DATA
HONEYWELL INC
DIAGNOSTICS
| <-- | -->
The text string currently stored in memory will be
displayed.
If desired, edit the string (nineteen characters maximum)
using the UP and DOWN keys, and the <-- and --> function
keys.
Press ENTER or MENU to exit.