Honeywell 9782 Series Network Card User Manual


 
I/O Setup and System Configuration
7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 5-9
Step Screen Action
13
SOLUTION TEMP COMP
C1 ACID/NH3/ETA
C2 NACL
DIAGNOSTICS
| PREV | NEXT
Select the type of solution for which temperature
compensation should be applied.
14
ANALOG OUTPUTS
OUTPUT1 C1
OUTPUT2 C2
OUTPUT3 RATO
DIAGNOSTICS
| PREV | NEXT
If the hardware supports analog outputs, specify for each
which process variable or computed value will be
associated with the output.
3
15
OUTPUT SCALING
OUTPUT1 LINFIXEDRNG
OUTPUT2 LINAUTORNG
OUTPUT3 BI-LINEAR
DIAGNOSTICS
| PREV | NEXT
If analog outputs will be used, specify the scaling for each
output.
4
CAT (Current Adjusting Type) control uses Fixed
Linear Range only. Retransmitted process values can
be scaled. A retransmitted computed value always
uses Fixed Linear Range.
16
I/O SETUP COMPLETE
PRESS
ENTER TO CONTINUE
DIAGNOSTICS
| PREV |
At this point, you have completed the I/O setup menu. If
you have changed at least one parameter (other than
solution temperature compensation) the unit will perform a
cold reset (see page 5-1).
UPDATING EEPROM
DIAGNOSTICS
| |
UPDATING EEPROM
PRESS ENTER
TO RETURN TO
MAIN MENU
DIAGNOSTICS
|
|
When you press ENTER to continue, the message
“UPDATING EEPROM” will provide feedback, letting you
know your selections have been activated.
When the update has been completed, you will be
prompted to exit the menu.
3
If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show
“DE ENABLED”, instead of displaying output assignments.
4
This output range scaling has no effect on the display. Display resolution is configured using the maintenance
menu as described in Section 10.