Honeywell 9782 Series Network Card User Manual


 
Unpacking, Preparation, and Mounting
7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual
3-3
Honeywell
York, PA U.S.A.
ATTENTION - UNIT SET FOR 240 VAC
9782C-S0-VC-E0010-00
S/N: 9751Y712345670001
USA WARRANTY ASSISTANCE - 1 - 800-423-9883
120/240 VAC 50/60 Hz 15VA
Contact Rating 1 & 2: 0.6 Amps at 120/240 VAC,
0.6 Amps at 110 VDC, and 2 Amps at 30 VDC.
If installed 3 & 4: 3 Amps at 120 VAC, 28 VDC.
CE
CASE: NEMA 4X Directions: 70-82-25-74
Figure 3-1 Sample Nameplate
3.3 Mounting
Introduction
The 9782 Analyzer/Controller can be panel-, wall- or pipe-mounted.
Panel-mounting
Figures 3-2-1 and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit) show panel cutout
dimensions. They also illustrate how the instrument is attached to a panel by clamping the edges
of the cutout between the case flange and the supplied U-bracket which is fastened to the rear of
the case.
The panel may be up to 3/ 8 in. thick. Cutouts for adjacent 9782 Analyzer/Controllers may be no
closer than .687 in. horizontally and .937 in. vertically, resulting in a nominal edge-to-edge bezel
separation of .078 in.
Note that if the panel already has a cutout for a 7082 instrument, the same cutout can be used for
the 9782. If the panel has a cutout for a 7070, 7075, 7076, 7077, or 7078 instrument, a Reducer
Panel Kit is available for mounting the 9782. The part number is in Section 11.
Wall- and pipe-mounting
A bracket is available as an option for wall or pipe mounting the unit as shown in Figures 3-2-1
and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit).
The mounting bracket part number is in Section 11.
ATTENTION
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration
may affect system performance.