Honeywell 9782 Series Network Card User Manual


 
I/O Setup and System Configuration
7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 5-17
5.7 Setting the Clock
Purpose
A display showing the current time can be accessed when the unit is online. The 24-hour clock
must be set before you can use the AutoClean feature described in Sub-Section 5.6.
Procedure
Follow the procedure in Table 5-7 to set the 24-hour clock.
Table 5-7 Procedure for Setting the Clock
Step Screen Action
1
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
| PREV | NEXT
Go to CONFIGURATION | ADVANCED FEATURES |
CLOCK SETUP.
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
| PREV | NEXT
ADVANCED FEATURES
CLOCK SETUP
COMPUTED VALUE
AUTO CLEAN
DIAGNOSTICS
| |
2
24 HOUR CLOCK SETUP
CLOCK OFF
HOURS 13
MINUTES 30
DIAGNOSTICS
| | NEXT
Set the clock. For example, if you will start the clock at
1:30 PM, enter 13 hours, 30 minutes.
Start the clock and specify what should happen to the
clock if the power fails.
ON W/RESET starts the clock now. Following power
failure, the clock will be reset to 00:00 and the diagnostic
message “RESET CLOCK” will be displayed.
ON W/O RESET starts the clock now. Following power
failure the clock will remain off, and the diagnostic
message “RESET CLOCK” will be displayed.
Display of ”RESET CLOCK” will continue until you either
reset the clock or turn it off.