Honeywell 9782 Series Network Card User Manual


 
Conductivity/Resistivity Analyzer/Controller
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/9910-14
10.9 Replacing the Fuse
Introduction
Figure 10-1 shows the location of the power fuse on the power supply card in the right-most
position inside the case. (See Section 11 for the part number of the kit containing fuses.)
9
8
7
6
5
4
3
Figure 10-2 Location of Power Fuse
WARNING
Disconnect power before opening the instrument case to replace the fuse. A potentially lethal
shock hazard exists inside the case if the unit is opened while powered. More than one switch
may be required to disconnect power.
ATTENTION
To avoid potentially serious damage to expensive components, a blown fuse should not be
replaced until the cause has been determined by a qualified service technician.
Procedure
To replace the fuse follow the procedure in Table 10-8.