9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/997-2
7.2 AutoClean Sequence and Piping
Rinse sequence
The AutoClean operation occurs at the configured intervals. (It can also be initiated by the
operator; see Sub-Section 5.8 for details.) The sequence is described below.
1 All alarm action is held at existing levels. The outputs are also held throughout the
AutoClean operation.
Pressing the DISPLAY key will call up a special display that shows how much time
remains until the start of the next scheduled AutoClean sequence. While the cleaning
operation is taking place, “AUTO CLEAN” is displayed on the alarm stripe on the Time
Remaining display. The clock will show 00:00:00.
2 Relay 1 activates 3-way solenoid valve S1 (see Figure 7-1) to direct rinsing fluid to the
cells for the configured rinse duration (1 to 1999 seconds). If the measured sample is
normally returned to the process but quantities of rinsing fluid cannot be tolerated there,
use an additional 3-way solenoid valve S4. It is activated simultaneously with S1 to divert
the discharge to drain.
3 At the end of the configured rinse time Relay 1 de-activates solenoid valve S1 (and S4, if
used). After the configured delay period (1 to 1999 seconds) the Analyzer/Controller
resumes sampling the process. (During the resume delay the “AUTO CLEAN” message
flashes.) Note that even with S4 for diversion, one system volume of washing fluid will
pass to the process at this point.
4 The “AUTO CLEAN” message is cleared.
Note that the operator can make the operation pause using the “soft key” on the special
AutoClean Time Remaining display. If the operator does not remove the pause by pressing the
PAUSE soft key again, the Analyzer/Controller will resume normal operation after 20 minutes.