Honeywell 9782 Series Network Card User Manual


 
Conductivity/Resistivity Analyzer/Controller
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/995-20
Step Screen Action
3
AUTO CLEAN
TIMER SETUP
DAYS 3
HOURS 16
MINUTES 30
CURRENT DAY 2
DIAGNOSTICS
| | NEXT
Timer setup is used to specify the frequency of the rinses.
If several rinses are to be performed in a single day:
a) specify the number of days as zero
b) specify the number of hours and minutes in the
interval between rinses.
If rinses are to occur at intervals of days:
a) specify the number of days in the interval
b) specify the time of day (hours and minutes) the
rinse should occur.
In our example, the rinse will occur every third day at
4:30 PM.
If you plan to use AutoClean, but initiate rinses manually
using the real time AutoClean Time Remaining display:
a) specify the number of days as zero
b) specify the number of hours and minutes in the
interval between rinses as zero also.
IF YOU ENTER ALL ZEROS, NO AUTOMATIC
RINSES WILL BE PERFORMED.
Specify where today falls in the AutoClean cycle. For
example, if the interval for rinses is 3 days and you want
the first rinse to be tomorrow, specify that today is day 2.
4
AUTO CLEAN
RINSE SETUP
DURATION 180
RESUME DELAY 60
DIAGNOSTICS
| PREV | NEXT
Enter the duration of the rinse (in seconds).
Enter the number of seconds the Analyzer/Controller
should wait after the rinse before resuming normal
operation.