Honeywell 9782 Series Network Card User Manual


 
I/O Setup and System Configuration
7/99 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 5-25
PARAMETER CHOICES or RANGE
(Default is underlined)
configuration: retransmission 0%
configuration: retransmission 100%
range depends on measurement type and cell constant;
default is value of lower range limit
configuration: CAT control setpoint value
configuration: alarm setpoint value
configuration: configuration: On/Off control cycle timer
setpoint value
configuration: PFT control setpoint value
configuration: DAT control setpoint value
range depends on measurement type and cell constant;
default is value of upper range limit
configuration: CAT control proportional band limit
configuration: alarm deadband
range depends on measurement type and cell constant;
default is value of upper range limit
configuration: On/Off control cycle timer deadband
configuration: PFT control proportional band limit
configuration: DAT control proportional band limit
configuration: alarm delay
0 to 100 seconds, default = 0
configuration: process variable or computed value to
be used as basis for discrete control output
configuration: process variable or computed value
alarmed
C1 [Cell 1 value]
C2 [Cell 2 value]
TMP1 [temperature value from Cell 1]
TMP2 [temperature value from Cell 2]
T1/2 [temperature 1 and/or 2 - only available for
alarming]
RATO [computed ratio:
Cell
Cell
1
2
]
%REJ [computed percent rejection:
()x1
1
2
100
Cell
Cell
]
%PSG [computed percent passage:
Cell
Cell
x
1
2
100
]
DIFF [computed difference:
Cell Cell12
]
CO2 [computed parts per million carbon dioxide -
9782C-S0 only]
configuration: alarm action
HIGH
LOW
configuration: On/Off control cycle timer period
0 to 1000 seconds, default = 60
configuration: On/Off control cycle timer “on” time
0 to 100% at 1% resolution, default = 100
configuration: PFT control maximum pulse frequency
1 to 200 pulses per minute, default = 60
configuration: DAT control cycle period
0 to 200 seconds, default = 60
configuration: 24-hour clock set mode
OFF
ON W/ RESET
ON W/O RESET
configuration: 24-hour clock set hour
0 to 24, default = 1
configuration: 24-hour clock set minute
0 to 59, default = 0
configuration: AutoClean timer set day
0 to 28, default = 0
configuration: AutoClean timer set hour
0 to 24, default = 1
configuration: AutoClean timer set minutes
0 to 59, default = 0