9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual 7/995-26
PARAMETER CHOICES or RANGE
(Default is underlined)
configuration: AutoClean current day
0 to 28, default = 0
configuration: AutoClean rinse duration
configuration: AutoClean resume delay
1 to 1999 seconds, default = 60
configuration: computed value selection
NONE
RATIO [computed ratio:
Cell
Cell
1
2
]
DIFFERENCE [computed difference:
Cell Cell12
]
%PASSAGE [computed percent passage:
Cell
Cell
x
1
2
100
]
%REJECTION [computed percent rejection
:
()x1
1
2
100
Cell
Cell
]
PPMCO2 [computed parts per million carbon dioxide -
9782C-S0 only]
maintenance: display resolution
LOW
HIGH
maintenance: security enabled
YES
NO
maintenance: password
0000 to 9999; default = 0000
maintenance: lead resistance compensation wire gauge
(AWG)
wide range models only
12
14
16
18
maintenance: lead resistance compensation length
wide range models only
1 to 1500 feet default = 1
calibration: TDS conversion factor
standard range models with cell constant of 1, 10, 25, or
50:
0.01 to 1.999; default = 0.5 ppm
standard range models with cell constant of 0.01 or 0.1:
10 to 1999; default = 500 ppb
wide range models for all cell constants:
0.01 to 1.999; default = 0.5 ppm