Xerox CX240 All in One Printer User Manual


 
Calibrating with the X-Rite DTP34 QuickCal Densitometer 113
Guidelines for Successful Calibration
To ensure your calibration is as accurate as possible, check the following
guidelines before you calibrate:
Ensure that your X-Rite DTP34 QuickCal densitometer is connected
properly
Ensure that your X-Rite DTP34 QuickCal densitometer is calibrated.
You should calibrate the device at least once a week, or whenever the
power to the device is interrupted. Use the manufacturer calibration
chart supplied with the device. After you finish calibrating the device,
make sure you store the chart in the designated envelope.
To warm up the printer, use any media to print at least 25 duplex sheets
of any four color test job.
Print a reference job and use the same media and screen type on which
you will print the final job.
Prepare the calibration pad to measure the calibration charts. If a
calibration pad is not available, set a white media, at least 200 gsm, or
two sheets of lighter paper, on the surface. Read the calibration chart
on this surface.
Set the Spire CX250 color server emulation method to the method you
will use to print the customer job (Device Link).
Calibrating the X-Rite DTP34 QuickCal Densitometer
The X-Rite DTP34 QuickCal densitometer is a color measurement
instrument that reports densitometer and dot data.
Perform the following steps before first time use of the X-Rite DTP34
QuickCal densitometer:
Familiarize yourself with the X-Rite DTP34 QuickCal Densitometer
Operations Manual.
Connect the X-Rite DTP34 QuickCal densitometer.
Calibrate the X-Rite DTP34 QuickCal densitometer.
You can also use the DTP34 Tutorial to learn how to use the X-Rite DTP34
QuickCal densitometer correctly. Access the tutorial from
Tools>Calibration, and then click Calibrate. In Step 1 of the Color
Calibration wizard, click DTP34 Tutorial - Click & Learn.