Agilent Technologies Switch/Measure Switch User Manual


 
Dual/Triple Microwave Switch Modules 8
34980A User’s Guide 183
Example: Querying the system for module identify The following command
returns the identify of the module installed in slot 7.
SYSTem:CTYPe? 7
Example: Reading the cycle count for a relay The following command reads
back the number of completed cycles for the channel 201 relay of a module
installed in slot 6.
DIAGnostic:RELay:CYCLes? (@6201)
Example: Clearing the cycle count for a relay The following command resets
the cycle count on the channels 201 and 202 for a module in slot 1.
DIAGnostic:RELay:CYCLes:CLEar (@1201,1202)
Example: Resetting Module(s) to Power-On State The following command
resets a module in slot 4 to its power-on state.
SYSTem:CPON 4
Installing SMA Connectors
When installing SMA connectors, it is recommend that you tighten them to
0.8 - 1.1 Nm (7-10 in-lbs) of torque.