Agilent Technologies Switch/Measure Switch User Manual


 
26 34980A User’s Guide
2 Features and Functions
Front Panel Operation: Channel (Configure) > CHANNEL LABEL
To define the channel label, press the arrow keys to move the cursor to a
specific position and then turn the knob to select the desired letter or
number.
To clear the channel label on the selected channel, change each character
to “ ^” (starting with the rightmost character) and then press the left
arrow key to move to the next character.
To clear all channel labels on the selected module, navigate to:
Module (Configure) > CLEAR LABELS? > YES
Remote Interface Operation: The following command assigns a label
(“TEST_PT_1”) to channel 3 in slot 1.
ROUT:CHAN:LABEL "TEST_PT_1",(@1003)
The following command clears the user- defined label previously assigned
to channel 3 in slot 1. The channel will now be identified by its factory
default label (e.g., “MUX CH BANK 1”, “MATRIX1 ROW3 COL4”,
“DIO BYTE 1”, etc.).
ROUT:CHAN:LABEL "",(@1003)
The following command clears all user- defined channel labels on the
module in slot 1. The factory-default labels are assigned to all channels on
the module in slot 1.
ROUT:CHAN:LABEL:CLEAR:MOD 1
The following command clears all user- defined labels on all modules
installed in the 34980A. The factory-default labels are assigned to all
channels on all installed modules.
ROUT:CHAN:LABEL:CLEAR:MOD ALL