Agilent Technologies Switch/Measure Switch User Manual


 
Features and Functions 2
34980A User’s Guide 57
Remote Interface Operation: Use the following commands to select the
reading format.
FORMat:READing:ALARm ON
FORMat:READing:CHANnel ON
FORMat:READing:TIME ON
FORMat:READing:TIME:TYPE {ABSolute|RELative}
FORMat:READing:UNIT ON
Non-Sequential Scanning
By default, the instrument scans the list of channels in ascending order
from slot 1 through slot 8 (channels are reordered as needed). If your
application requires non-ordered scanning of the channels in the present
scan list, you can use the non- sequential scanning mode.
This feature is available from the remote interface only.
The scanning mode applies to the entire mainframe and cannot be
selectively used on individual modules.
When sequential scanning is enabled (default), the channels in the scan
list are placed in ascending order from slot 1 through slot 8. Duplicate
channels are not allowed. For example, (@2001,1003,1001,1003) will be
interpreted as (@1001,1003,2001).
When sequential scanning is disabled (OFF), the channels remain in the
order presented in the scan list (see exception below). Multiple
occurrences of the same channel are allowed. For example,
(@2001,2001,2001) and (@3010,1003,1001,1005) are valid and the
channels will be scanned in the order presented.
When you specify a range of channels in the scan list, the channels are
always sorted in ascending order, regardless of the scan order setting.
Therefore, (@1009:1001) will always be interpreted as 1001, 1002,
1003, etc.
If you define a scan list with the sequential mode enabled and later
disable the mode, the scan list will not be reordered; however, the scan
list will be treated as a non-sequential list thereafter.
If you have defined a scan list with the sequential mode disabled (OFF)
and later enable the mode, the channels will be reordered.
Non-sequential scan lists are not stored as part of the instrument state
by the
*SAV command; in this case, the ordered mode will be enabled
and the scan list will be empty when the instrument state is restored
(*RCL command).