The CONFigure and MEASure? commands automatically set the gain
(“M”) to 1 and offset (“B”) to 0.
A Factory Reset (
*RST command) turns off scaling and clears the
scaling values on all channels. An Instrument Preset (SYSTem:PRESet
command) does not clear the scaling values and does not turn off scaling.
Front Panel Operation:
DMM or Channel (Configure) > SCALING > GAIN|OFFSET|UNITS
To define the label on the selected channel, press the arrow keys to move
the cursor to a specific position and then turn the knob to select the
desired letter or number. To clear the label on the selected channel,
change each character to “ ^” (starting with the rightmost character) and
then press the left arrow key to move to the next character.
Remote Interface Operation: Use the following commands to set the gain,
offset, and custom measurement label.
CALC:SCALE:GAIN 1.2,(@1003)
CALC:SCALE:OFFSET 10,(@1003)
CALC:SCALE:UNIT 'PSI',(@1003)
After setting the gain and offset values, send the following command to
enable the scaling function on the specified channel.
CALC:SCALE:STATE ON,(@1003)