Agilent Technologies Switch/Measure Switch User Manual


 
48 34980A User’s Guide
2 Features and Functions
Remote Interface Operation: The following program segment configures the
instrument for a manual scanning operation.
Then, send the
*TRG (trigger) command to begin each scan sweep.
The
*TRG command will not be accepted unless the internal DMM is in
the “wait-for-trigger” state.
Note: To stop a scan, press and hold the Scan (Measure) key.
External Scanning
In this configuration, the instrument sweeps through the scan list once
each time a low- going TTL pulse is received on the rear-panel Ext Trig
Input line (pin 6).
You can specify a scan count which sets the number of external pulses
the instrument will accept before terminating the scan. See “Trigger
Count” on page 49 for more information.
If the instrument receives an external trigger before it is ready to
accept one, it will buffer one trigger and then ignore any additional
triggers received (no error is generated).
All readings from the scan are stored in non-volatile memory.
Readings accumulate in memory until the scan is terminated (until the
scan count is reached or until you abort the scan).
Mx+B scaling and alarm limits are applied to measurements during the
scan and all data is stored in non-volatile memory.
TRIG:SOURCE BUS
TRIG:COUNT 2
INIT
Select bus (manual) mode
Sweep the scan list 2 times
Initiate the scan
Ext Trig Input connector (as viewed from rear of instrument)
Input
5 V
0 V
> 1 µs
Ext Trig Input (Pin 6)
Gnd (Pin 9)
1
5
6
9